Changes

Jump to navigation Jump to search
Line 19: Line 19:  
=== Proposed Measurement Program ===
 
=== Proposed Measurement Program ===
   −
 
+
# Irradiation with about 40 rem over the course of a few days (source strength selected appropriately)
 +
## periodically acquire an ensemble of pulses in the dark to determine gain, and dark rate
 +
## periodically acquire an ensemble of pulses in the with a pulser to determine gain, and photo-detection efficiency
 +
# Relaxation time of duration determined by the apparent healing rate.
 +
## periodically acquire an ensemble of pulses in the dark to determine gain, and dark rate
 +
## periodically acquire an ensemble of pulses in the with a pulser to determine gain, and photo-detection efficiency
 +
# Repetition of 1. and 2. as desired to determine if complete healing is possible at these doses
 +
# Irradiate the unit for a much longer duration - until some saturation of the performance parameters is seen. This will help determine the offset in the model of the deterioration of performance parameters.
 +
## periodically acquire an ensemble of pulses in the dark to determine gain, and dark rate
 +
## periodically acquire an ensemble of pulses in the with a pulser to determine gain, and photo-detection efficiency
    
== Links ==
 
== Links ==
1,004

edits

Navigation menu