|+ Summary of SiPM Properties measured at room temperature at recommended bias voltages
+
|+ ''Summary of SiPM Properties measured at room temperature at recommended bias voltages''
|- style="background:#F1F2FF"
|- style="background:#F1F2FF"
!rowspan="2"| Device
!rowspan="2"| Device
Line 92:
Line 92:
|}
|}
+
Since the SiPM performance is sensitive to the bias voltage applied and the ambient temperature, a measurement SiPM properties as functions of bias voltage (V<sub>b</sub>) and temperature (T) was performed on the SSPM-06~. (By this point, the SSPM-06~ was judged to a better sensor for the tagger microscope, owing to higher sensitivity in the blue-green range and better active area match to the fiber cross-section. Aside from gains in efficiency and dynamic range of the resulting readout, higher photon detection implies higher time resolution because of the scintillation decay time of 2.7ns in the fiber.)
+
+
The range of interest for these operating variables were:
+
- V<sub>b</sub>: from 0.5V below to 0.5V above the operating range: 19V-21V
+
- T: 0-above room temp., in practice 3°C (to avoid growing snow) to 25°C
+
+
However, it was found that the peaks were very indistinct by at bias voltages below 20V and temperatures above 20°C. This