LGD calibration logbook

Richard Jones
Jeff Gunter
David Armstrong
Mihajlo Kornicer
started May 2, 2000
last updated May 10, 2001

 
Table of Contents
  1. Was the electron beam energy 4.4GeV instead of 5.5GeV summer 1999?
  2. How is the shower depth in the LGD affected by attenuation?
  3. How do shower depth fluctuations get passed to the kinematic fitters?
  4. How is the shower depth in the LGD affected by wrapping?
  5. Does the new angle-dependent energy correction have a mass/energy bias?
     

 

Was the electron beam energy 4.4GeV instead of 5.5GeV summer 1999?

April 27, 2000 S. Teige issued an internal note [1] describing a surprising result that he came across while studying the ω sample from the July, 2000 run. Using the LGD to give the total momentum of the forward system, and using minimum - missing - momentum to estimate the momentum of the recoil proton, he was able to reconstruct the incident photon energy and compare it to what was seen in the tagger. The surprise was that the beam energies he obtained using this method came out so low that not only did they disagree with the tagger values, but they were outside the nominal tagging range. Scott concludes his note with the following request for comments, to which I reply below.


[ST] April 27, 2000
[RTJ] May 8, 2000
[RTJ] May 9, 2000
[JJK] May 9, 2000
[RTJ] May 10, 2000

 

How is the shower depth in the LGD affected by attenuation?

January 11, 2001 R. Jones wrote a note [2] recommending a modified formula for calculating the average depth of a shower in the LGD. This depth is needed, together with the shower position in (x,y), to determine the direction of the photon momentum vector. In that study it was assumed that the Cerenkov yield is proportional to the length of charged tracks in the glass. It was pointed out that attenuation of the Cerenkov light on the way to the phototube might produce important modifications to these results.

[JG] January 14, 2001

    Still lurking on the radphi list, I noticed your note on shower depths. Good stuff ... makes me wonder if geant has been improved in the decade since serious glass simulations were attempted for E852. Here are a couple of comments from the peanut gallery:

  1. Are your results stable w.r.t. changes in the minimum electron energy. If I recall correctly, some of the Notre Dame guys in E852 did a similar study but with energies down to about 5 or 10 keV... I seem to remember that it mattered a little.
  2. The effective depth can appear deeper than in your simulation because cerenk photons near the up stream end of are attenuated ... also if I recall correctly, not so much by the glass itself but by the loss at reflections as they bounce down the block.

[RTJ] January 17, 2001

    Jeff, thanks for keeping an eye on this stuff. On your first point, it depends on what you are measuring how sensitive the results are to the tracking cutoff. Below about 400keV the electrons stop making light. It turns out that the range of a 400keV electron in lead glass is about 50 microns, so cutting off there essentially has no effect on the track length. This does not prove that Cerenkov yield and track length are proportional, but at least neither one are affected by the tracking cutoffs in GEANT. On your second point, my feeling is that attenuation will not shift the depth centroid by an appreciable fraction of the r.m.s. fluctuations. But since I cannot come up with a better argument than that, I will do a study. I take the following features for my attenuation model.

  1. On every e+/e- track in the shower generate Cerenkov photons according to the standard formula, where the refractive index in general depends on the energy k of the Cerenkov photon.
    (7)
    where the numerical value of the constant No is 370 photons/cm/eV.
  2. Generate Cerenkov photons within a fixed transmission window between 800nm and 300nm, bounded above by the photocathode quantum efficiency and below by the opacity of lead glass.
  3. Generate Cerenkov photons with the correct polarization given by Cerenkov's theory and correct the polarization information at each reflection/refraction.
  4. Propagate the photons with exponential attenuation given by the attenuation curve of lead glass. At each boundary calculate the T and R coefficients and do Monte Carlo transmission/reflection.
  5. Any photons which transmit into the front surface of the phototube are detected according to the spectral quantum efficiency of the cathode.

    My attenuation length for lead glass is shown in Fig. 2, which I calculated based upon some transmission measurements that I received from S. Teige under the assumption that the refractive index of lead glass is 1.62. For the photocathode quantum efficiency, I just took a sample spectrum from Hamamatsu's web site for a bialkali cathode. This is shown in Fig. 3. A single shower of 1GeV generates of order 1 million Cerenkov photons, each of which have to be tracked by GEANT. This turns out to be quite computationally expensive so I only generated a few hundred showers. Nevertheless the results are sufficiently precise to answer the question. To avoid having to guess the spectral reflectivity of aluminized mylar, I decided to simply wrap the glass in black paper for this simulation. By this I mean that outside the glass is a material with unit refractive index and zero reflectance. The typical photon undergoes several reflections before absorption, so this is not the same as painting the surface black.

    The color spectrum of the detected Cerenkov photons is shown in Fig. 4. The spectrum is zero at the edges of the window, showing that the choice of window was OK. In Fig. 5 is plotted the originating z coordinate for every detected photon in a set of 100 showers from 1GeV gamma rays at normal incidence. This result will not be sensitive to the exact angle of incidence because the shower tracks are disoriented in the bulk of the shower. The average z-origin of detected Cerenkov photons is within a few mm of the depth predicted in Ref. [2]. I conclude that the effective shower depth after attenuation is taken into account does not differ appreciably from the physical one.

How do shower depth fluctuations get passed to the kinematic fitters?

[JG] January 19, 2001

    Since radphi is getting into the kinematic fitting business, one thing that I don't think people have worried about is the stat uncertainty in depth correction from shower to shower. In E852 the depth correction was a few percent of the target to glass distance (iirc about 20cm/535cm) and people never really worried with including fluctuations in interaction to shower distance in fits ... especially for all neutral final states with no good vertex information anyway ... target to lgd distance always had a sigma of order ~10cm (30 inch LH_2 target). This could be different radphi since target-lgd distance is much smaller and the target position is well known, even though its momentum isn't. I don't have a feel for this, but what kinds of shower to shower depth fluctuations in the depth are there? If they are large (? what's that mean ?) then the x and y uncertainties should probably be increased before the kinematic fit.

[RTJ] January 22, 2001

    If you remember from the second figure in my note, the r.m.s. depth fluctuations for a 1GeV shower are 4.3cm. The r.m.s. depth of the initial gamma conversion is 9/7 * X0 = 4.0cm so you can see that the fluctuations are essentially given by where the first pair in the shower originates. And since the radiation length is not a strong function of energy above 100MeV we can take this number 4.3cm to be a constant.

    At 20° this translates to an error in x,y from depth fluctuations alone, of 1.5cm. This is larger than the error from shower shape flucts, which was about 1cm r.m.s. from E852 if I remember my earlier work correctly. This may require us to use an elliptical error matrix on the (x,y) for each shower in the kinematic fits. That is not hard to do, but requires more knobs that we have to know how to control. I think that before we go that far we should have a study that shows observational evidence for off-diagonal error matrix elements in (x,y) for individual showers. Brent Evans is working with me on a study to look for this effect.

How is the shower depth in the LGD affected by wrapping?

Under item 2 above, a detailed simulation was described that followed Cerenkov photons from their source in the shower down the length of the block to the phototube. For simplicity, the simulation had assumed that the blocks are surrounded by a layer of air and that any photons escaping from the block are absorbed. It had also no air gap between the phototube and the glass at the end of the block. Under these conditions it was seen that the effective shower depth centroid is surprisingly insensitive to attenuation in the glass, shifting only a few mm when attenuation is turned on and off. This is small compared to the scale of shower depth fluctuations of about 4.5cm r.m.s.

[DA] January 29, 2001

    That result may depend on the way the glass is wrapped. I have seen that wrapping scintillator in aluminized mylar can make a paddle very sensitive to light emitted near the end close to the tube. This would not be exponential, and it might produce a larger effect than what you are seeing with bulk attenuation.

[RTJ] January 30, 2001

    Yes there may be an effect due to wrapping. To see this I will revise the model described above to remove the air gap and replace it with aluminum. I consider the thin layer of mylar to be irrelevant because of the small difference of its refractive index with that of lead glass. I take the reflectivity of aluminum to be 90% at all wavelengths and incidence angles. I also insert a 5mm air gap between the back end of the lead glass block and the phototube. There is a noticable effect from the wrapping! This can be seen in Fig. 6. The open histogram is the profile of the generated light, rescaled for comparison to the profile of light that is eventually detected at the phototube. As David said, there is an enhancement in the sensitivity to light generated near the downstream end of the block that comes from the wrapping. As a consequence, there is a shift of the depth centroid downstream by about 2cm.

    Fig. 7 shows the profile of the shower depth centroid vs incident gamma energy. The error bars show the r.m.s. fluctuations from shower to shower. These showers have been generated uniformly over the face of the block to avoid any bias from the alignment of the shower with the phototube. The black curve in Fig. 7 indicates Eq. 3 in [2]. The red curve has added 2cm.

Does the new angle-dependent energy correction have a mass/energy bias?

In Ref. [4] is described a procedure that introduces an angle-dependence to the conversion from shower pulse-heights to total shower energy. This is necessary in the Radphi forward calorimeter because at angles beyond 20° a rapidly increasing fraction of the shower energy leaks out the sides of the detector. In the analyses before this study was performed, it was understood that a simple linear model of pulse-heights vs shower energy was inadequate, and would produce a bias towards higher masses for decaying particles with larger energy in the lab frame. A nonlinear correction was used to cancel out this bias, by making the gain constant proportional to some small power (called epsilon in the code) of the total pulse height. By trial and error D. Armstrong found that the systematic shift of the π° peak position with π° lab energy could be approximately nulled out with a value of =-0.06 using the default calibration constants that have been in place since the summer 2000 run. Now that this nonlinearity correction has been superceded by the new angle-dependent procedure, the question arises whether the new algorithm exhibits any mass/energy bias.

[DA] March 9, 2001

    Initial results from testing lgdtune, Richard's new calibration code. Based on run 8342, 10 iterations, 500K events. Starting point for iterations was gain constants from the database (therefore based on older non-linearity correction code). The following test was done to look for walk of the π° peak position with total cluster multiplicity (comment: for constant total energy in the LGD, average photon energy is inversely proportional to the total multiplicity). The total energy in the LGD was required to be at least 4.0GeV. Any pairs were counted as a π° if their invariant mass falls within 20MeV of the physical value. The resulting mass spectra were fitted to a Gaussian plus a cubic background. The sigmas (widths) appear significantly better than with the previous calibrator; however the energy cut was different so the comparison may not be fair. The following fits were made.

    1. all events: centroid = 134.3MeV
    2. 2-cluster events: centroid = 141.2MeV
    3. 3-cluster events: centroid = 133.8MeV
    4. 4-cluster events: centroid = 131.0MeV
    The trend to lower mass centroids continues for 5 and 6 cluster events.
[RTJ] April 16, 2001

    This result of Dave's is surprising, since comparisons of the energy curves for fixed angle in the region below 10° show good agreement between the new procedure and the old formula with =-0.06. This is shown in Fig. 8. Between 10° and 20° the shape remains the same, but the gains move to lower values, reflecting the fact that at larger angles the shower leakage decreases by as much as 10%. This kind of systematic shift in gain with angle will just be absorbed into the gains of individual blocks, and so is not observable if the calibration is done with the same procedure as the analysis. The comparison between the old epsilon treatment and the new procedure for angles between 10° and 20° is shown in Fig. 9. At least the showers within the forward 20° cone should show the same mass/energy bias as with the old treatment, and for the clusters beyond 20° the new treatment should be a significant improvement. At UConn we will try to reproduce the effect seen by D. Armstrong.

[MK,RTJ] April 23, 2001

    The following test was performed on the data from run 8600. First Richard ran lgdtune, with the angle-dependent depth correction in place, to set new calibration constants for the counters. These calibration constants were then used to generate a sample of π°s. We were careful to use the same makehits calls for the analysis as were used for the lgdtune calibration (with lgd_cluster_cleanup and a total energy cut at 4.5GeV). The calibration was based only on π°s from 3-cluster events, but in the analysis /π°s from all events were collected and mass spectra were fitted. The centroids are given below:

    1. 2-cluster events: centroid = 142.7 MeV
    2. 3-cluster events: centroid = 134.8 MeV
    3. 4-cluster events: centroid = 132.4 MeV
    4. 5-cluster events: centroid = 130.7 MeV
    5. 6-cluster events: centroid = 130.9 MeV

    If our interpretation is correct, it is not the topology of these events that is shifting the mass but merely that the average energy of the two photons is anticorrelated to cluster multiplicity. If this is correct then the same effect should be visible by selecting only 2-cluster events and binning them in total energy. The following results taken from the same analysis (2 clusters only) give evidence that this reasoning is correct.

    1. total energy < 3.5 GeV (average 3.11GeV): centroid = 135 MeV
    2. 3.5<total energy<4.5 GeV (average 3.88GeV): centroid = 142 MeV
    3. total energy > 4.5 GeV (average 4.80GeV): centroid = 147 MeV
[MK,RTJ] April 24, 2001
    Another more explicit way to see the dependence of the π° mass on the energy of the participating clusters is to recall the formula for the mass.
    (8)
    where is the opening angle between the two photons and the square-root factor is just the geometric mean of the two cluster energies, which I call E12. A plot of E12 vs should show a deviation from the iso-mass curve if there is a bias. This is shown in Fig. 10 where the red curve represents the locus of π° decays and the blue represents the η. The approximation of twice the sine of the half-angle by the angle is good to 1% out to =0.5. This plot shows the reason for the apparent bias towards higher π° masses as E12 increases. The reason is the sharp cutoff that is imposed on the minimum opening angle by the condition that clusters can only be resolved if they are a certain distance apart. This angle of 0.075 r that appears on the plot corresponds to about 10cm at the LGD, about the minimum separation distance required to resolve two clusters. This cutoff allows π°s in if they fluctuate outward in opening angle, but nixes them if they fluctuate inward. These outward fluctuations are identified with upward mass fluctuations. So the bias is just an artifact of the minimum cluster separation cut in the lgd. In fact, if it were not observed there would be a mystery. Apart from the region at highest E12 values, the π°s locus appears to follow the expected shape given by the red line.

    We conclude that the present calibration procedure is not producing these large mass shifts through any procedural error, but that they are the consequence of our detector acceptance. A confirmation of this can be seen in the above data tabulated for higher cluster multiplicies, where the typical E12 values are lower and opening angles tend to be larger than the acceptance cutoff. There one sees a convergence of the downward trend to a fixed value. The calibration should be trained to tune this fixed value to the physical mass of the π°. The lgdtune program has now been modified in this way, and results will be forthcoming shortly.

[MK,RTJ] May 7, 2001
    It is decided to calibrate LGD by using the sample of π° with separation angle above some threshold. The threshold was chosen so that converging value of π° is close to its physical value, and to obtain stable η at the same time. For the minimum openning angle 0.085 r the following π° and η centroids are found:

                                                    π°                   η
    1. 2-cluster events: 144.5 MeV;   548.5 ± 0.5 MeV
    2. 3-cluster events: 137.3 MeV;   549.9 ± 0.5 MeV
    3. 4-cluster events: 134.3 MeV;   552.0 ± 2.5 MeV
    4. 5-cluster events: 133.2 MeV
    5. 6-cluster events: 132.4 MeV
    6. 7-cluster events: 131.8 MeV

    The mass dependance on cluster multiplicity corresponds just to the dependance on average cluster energy, so we decided to check further the "slope" of η's iso-mass curve in the E12- plot, Fig. 11. We switched to hyperbolic coordinate system and looked η mass in areas tied by different iso-normals (called f ) of the η's iso-mass curve. The following table shows how η mass depends on f :

      f :     < 0.0     ( 0.0, 1.0)   ( 1.0, 2.0)   ( 2.0, 3.0)   ( 3.0, 4.0)   ( 4.0, 5.0)     > 5.0  
      η     [MeV]:   524.7   540.4   549.0   548.9   552.1   554.6   564.0

    The walk of η mass in the low and high f areas does not support conclusion that this is due LGD acceptance. Thus, there is still some room for tuning angle-dependent energy non-linearity corrections.

[MK,RTJ] May 15, 2001
    When nonlinearity exponent (see [4], Eq. 7) is turned from 1.11 to 1.15 following dependence of pi end eta masses on cluster multiplicity was obtained, as well as eta mass dependence on f-cut:

      cluster multiplicity   π° [MeV]   width   η     [MeV]:   width
      2   142.9 ± 0.02   18.3 ± 0.02   537.6 ± 0.08   38.2 ± 0.1
      3   136.4 ± 0.02   16.7 ± 0.02   541.6 ± 0.3   40.0 ± 0.4
      4   134.3 ± 0.02   16.4 ± 0.02   543.7 ± 2.1   37.4 ± 2.8
      5   133.8 ± 0.05   17.4 ± 0.06    
      6   133.7 ± 0.1   17.6 ± 0.2    
      7   133.5 ± 0.3   17.7 ± 0.5    

      f :   (-1.0, 0.0)   ( 0.0, 1.0)   ( 1.0, 2.0)   ( 2.0, 3.0)   ( 3.0, 4.0)   ( 4.0, 5.0)     > 5.0  
      η     [MeV]:   524.7   535.3   540.3   538.3   539.1   540.6   549.9

    Eta mass is shifted down, but more significant is that pi0 mass is stabilized around 134 MeV for higher cluster multiplicities, and f-dependence of eta mass shows less walk then before. We concluded that energy nonlinearity is not the major thing that has to be tuned (we got right pi/eta ration with previous nonlinearity) but rather depth correction. For higher energy Z is pushed backwards and that increases angle so eta lies above iso-mass line. For low energy hits Z is pushed forward and theta is turned smaller so eta walks down below eta-mass line.

[MK,RTJ] May 16, 2001
    Nonlinearity exponenet is returned to 1.11 and new terms in energy-angle dependant depth correction are introduced: Z = Z0 + Z1 - Z2. Z0 is as before, Z1 term includes logarithmic energy dependance while Z2 term is proportional to 4-th power of theta. The net result is that eta/pi0 mass ratio is restored but the pi0 and eta walks as well.

 

Useful links

[1] S. Teige, "Towards Constrained Fitting for RadPhi" (April, 2000).
[2] R.T. Jones, "Estimation of the average depth of showers in the LGD" (January, 2001).
[3] S. Teige, "A Plan for Calibration" (March, 1998).
[4] R.T. Jones, "Energy nonlinearity and depth correction of showers in the LGD" (February 2001).


This page is maintained by Richard Jones.
Last modified